OraChrom, Inc.

 
       

 

 

Home
Up

 

FYI NOTE

FYI041698

Resolving capabilities of a 25 cm column without the flow and pH restrictions of Silica.

Human Angiotensins with one to three amino acid variances can be readily separated with high resolutions even at very high speed by altering their retentions through pH adjustments.

 

    Angiotensin I:   Asp-Arg-Val-Tyr-Ile-His-Pro-Phe-His-Leu

    Angiotensin II:  Asp-Arg-Val-Tyr-Ile-His-Pro-Phe

    Angiotensin III:        Arg-Val-Tyr-Ile-His-Pro-Phe

 

 

Such separation is not possible with Silica under normal operating conditions:

 

The same separation could be achieved significantly faster by using the same polymeric column at higher flow rates:

Operating parameters:

HPLC System.

HP 1100, standard cell.

Column

As indicated on each chromatogram.

Mobile Phase

 

pH = 2

A: 0.1% TFA in H2O
B: 0.1% TFA in Acetonitrile/H2O (95/5)(v/v)

pH = 11.5

A:7 mM PO4Na3 in H2O
B: 7 mM PO4Na3 in ACN:H2O (50:50)

Gradient

Optimized in each case

Temperature

20°C.

Detection

214 nm

 

Polymeric Stationary Phases, Setting new standards in High Speed, High Performance,

Liquid Chromatography.

Home ] Up ]