Hydrophobic Interaction Chromatography: Comparison of STYROS HIC-Ether with TSKgel Ether-5PW from TOSOH.

 

A mixture of 7 proteins were separated on a STYROS™ HIC-Ether 4.6 x 100 mm column (volume 1.7 ml) at linear flow rates of 720 cm/hr
(2 ml/min volumetric flow) and compared with the performance of a TOSOH TSKgel Ether-5PW 7.5 mm x 7.5 cm, column (volume 3.3 ml) run by the manufacturer at a linear flow rate of 136 cm/hr (1 ml/min volumetric flow) using 5 proteins.

Both resins are porous; STYROS™ is a Simulated Monolith column made of hard gel polymeric that can withstand high back pressures and therefore can be run at high flow rates, whereas TSKgel is a soft gel resin with pressure and flow rates limitations.

 

Chromatogram 1

Separation of 7 proteins on STYROS HIC-Ether/XH

(Linear Flow Rate: 720 cm/hr)

 

Table 1. Operating parameters.

HPLC System.

Agilent 1100 with thermostatted column compartment.

Columns

STYROS HIC-Ether/XH 4.6 X 100 mm

Mobile phase.

A: 0.1 M Phosphate, pH=7
B: A + 2.1 M SO4(NH4)2, pH=7

Flow rate

2 ml/min (720 cm/hr )

Gradient

100 to 30 % B in 10 min (12 cv).

Temperature

30°C

Detection

280 nm

Injection volume

10 ml

Sample:

1- Cytochrome c, 0.1 mg/ml, 2- Myoglobin 2.5 mg/ml, 3-Ribonuclease A,
5 mg/ml, 4- Lysozyme 2 mg/ml, 5- Ovalbumin 5 mg/ml, 6-
α-Chymotrypsin 2.5 mg/ml, 7- α-Chymotrypsinogen A 0.5 mg/ml in buffer A.

 

Chromatogram 2

Separation of 5 proteins on TSKgel Ether-5PW

(Linear Flow Rate: 136 cm/hr)

 

Table 2. Operating parameters.

Columns

TSKgel Ether-5PW, (7.5 mm x 7.5 cm)

Mobile phase.

A: 0.1 M Phosphate, pH=7
B: A + 1.8 M SO4(NH4)2, pH=7

Flow rates

1 ml/min (136 cm/hr)

Gradient

100 to 0 % B in 60 min

Temperature

25°C

Detection

280 nm

Injection Volume

100 µl (50-100 µg)

Sample 1- Myoglobin, 2- Ribonuclease, 3- Lysozyme, 4- α-Chymotrypsin,
5- α-Chymotrypsinogen.

 

TSKgel Ether-5PW is a soft gel that operates at low back pressure (150 psi), with the maximum pressure being 300 psi.

STYROS™ media in comparison are made of fully pervious polymeric hard gel. They can withstand back pressures of up to 3000 psi.

The back pressure of the 4.6x100 mm column in the present application, run at 2 ml/min, is under 1500 psi at the start of the gradient with 2.1 M salt.

It is therefore possible to run separations at low flow rates with low back pressures for FPLC applications and also use longer STYROS™ columns (250mm) to explore added resolution.

Such versatility is very unique with this line of products.