APPLICATION NOTE

AN062506-41


Hydrophobic Interaction Chromatography : Comparison of STYROS HIC-Phenyl with Shodex HIC PH-818.

 

A mixture of 6 proteins were separated on a STYROS™ HIC-Phenyl 4.6 x 100 mm column (volume 1.7 ml) at linear flow rates of 720 cm/hr and 120 cm/hr and compared with the performance of a typical Shodex HIC Phenyl column of 8.0 mm x 75 mm (volume 3.8 ml) run at a linear flow rate of 120 cm/hr (1 ml/min volumetric flow).

The results show a clear advantage with the STYROS™ column both in performance and operating conditions.

 

Separation of 6 proteins on STYROS HIC-Phenyl/XH

(Linear Flow Rate: 720 cm/hr)

 

 

Chromatogram 2

Separation of 6 proteins on STYROS HIC-Phenyl/XH

(Linear Flow Rate: 120 cm/hr)

 

Table 1. Operating parameters for chromatograms 1 and 2.

HPLC System.

Agilent 1100 with thermostatted column compartment.

Columns

STYROS HIC-Phenyl/XH 4.6 X 100 mm

Mobile phase.

A: 0.1 M Phosphate, pH=7
B: A + 2.1 M SO4(NH4)2, pH=7

Flow rates

2 ml/min (720 cm/hr ) and 0.33 ml/min (120 cm/hr)

Gradient

100 to 0 % B in 9 min (11 cv) and 100 to 0% B in 55 min (11 cv)

Temperature

30°C

Detection

280 nm

Injection volume

10 ml

Sample:

1-Cytochrome c, 0.5 mg/ml, 2-Myoglobin 2.5 mg/ml,
3-Ribonuclease A 5 mg/ml, 4-Lysozyme 2 mg/ml, 5- Ovalbumin
5 mg/ml, 6- α-Chymotrypsinogen A 2.5 mg/ml in buffer A.

 

Chromatogram 3

Separation of 6 proteins on Shodex HIC PH-814

(Linear Flow Rate: 120 cm/hr)

Table 1. Operating parameters.

Columns

Shodex HIC PH-814 (8.0 mm x 75 mm)

Mobile phase.

A: 0.1 M Phosphate, pH=7
B: A + 2.1 M SO4(NH4)2, pH=7

Flow rates

2 ml/min (720 cm/hr ) and 0.33 ml/min (120 cm/hr)

Gradient

100 to 0 % B in 60 min (16 cv) and 100 to 0% B in 55 min (11 cv)

Temperature

30°C

Detection

280 nm

Injection volume

10 ml

Sample:

1-Cytochrome c, 0.03%, 2-Myoglobin 0.08%,
3-Ribonuclease A 0.16%, 4- Ovalbumin 0.16%, 5- Lysozyme 0.04%, 6- α-Chymotrypsinogen 0.05%.