APPLICATION NOTE

AN041006-40


Hydrophobic Interaction Chromatography compared with Polymeric Reversed Phase: STYROS HIC-Butyl versus STYROS 2R.

 

Hydrophobic Interaction Chromatography or HIC is based on the adsorption of biomolecules such as proteins through non ionic interactions between non polar regions on the protein’s surface and the hydrophobic surface of the stationary phase. It is usually performed during an elution starting with high salt concentrations.

The following chromatogram shows the fast separation of 6 proteins on a 10 cm column on HIC mode.

 

Separation of 6 proteins on STYROS HIC-Butyl/XH

 

Table 1. Operating parameters for the chromatograms.

HPLC System.

Agilent 1100 with thermostatted column compartment.

Columns

STYROS HIC-Butyl/XH 4.6 X 100 mm

Mobile phase.

A: 0.1 M Phosphate, pH=7
B: A + 2.1 M SO4(NH4)2, pH=7

Flow rate

2 ml/min (720 cm/hr )

Gradient

100 to 0 % B in 9 min (11 cv)

Temperature

30°C

Detection

280 nm

Injection volume

10 ml

Sample:

1-Cytochrome c, 0.5 mg/ml, 2-Myoglobin 2.5 mg/ml, 3-Ribonuclease A  5 mg/ml, 4-Lysozyme 2 mg/ml, 5-Ovalbumin 5 mg/ml, 6- α-Chymotrypsinogen A 2.5 mg/ml in buffer A.

 

HIC is used following a step that resulted in high salt concentration such as precipitation of the protein with ammonium sulfate, ion exchange chromatography or simply as the initial step from the salt containing biological medium.

Reversed phase chromatography or RPC, on the other hand consists of binding the proteins in a polar mobile phase and reducing the polarity of the mobile phase during elution.  

The reversed phase separation of the same mixture of proteins is shown in the following chromatogram with a similar size column.

Separation of 6 proteins on STYROS 2R/XH

 

Table 2. Operating parameters for the chromatogram.

HPLC System.

Agilent 1100 with thermostatted column compartment.

Columns

STYROS 2R/XH 4.6 X 100 mm

Mobile phase.

A: 0.075 % TFA in H2O.
B: 0.075 % TFA in ACN:H2O (95:5)

Flow rate

2 ml/min (720 cm/hr )

Gradient

30 to 50 % B in 9 min (11 cv)

Temperature

30°C

Detection

280 nm

Injection volume

10 ml

Sample:

1-Cytochrome c, 1 mg/ml, 2-Myoglobin 2.5 mg/ml, 3-Ribonuclease A 1.5 mg/ml, 4-Lysozyme 1 mg/ml, 5- Ovalbumin 5 mg/ml,
6- α-Chymotrypsinogen A 2.5 mg/ml in buffer A.

 

Comparison of the two methods:

HIC

RPC

Non denatured proteins

Denatured proteins

Adsorption chromatography

Partition chromatography

Weaker interaction

Stronger interaction

Less hydrophobic ligands

More hydrophobic ligands

Elution with reducing salt in water.

Elution with organic non polar solvents.

Matrix less substituted

Matrix more substituted

 

With STYROS™ columns both methods can be used in high and low pressure chromatography mode.