HPLC System. |
Agilent 1100 |
|
Columns |
StyrosZymeä
Papain 4.6 x 33 mm |
|
Mobile Phase For reversed phase. |
A: 0.1% TFA in H2O |
|
Digestion buffer. |
20 mM Phosphate, 20 mM Cysteine-HCl, 10 mM EDTA, at pH=6.2. |
|
Detection |
214 nm |
|
Gradient for Separation |
12 to 37 % B in 70 min @ 0.5 ml/min flow rate |
|
Injection volume |
8 ml |
|
Sample: |
10 mg/ml Cytochrome c from horse heart in 0.1 % TFA. |
By modifying the temperature of the StyrosZymeä Papain column, or changing the flow rate during the on line digestion, one can alter and control the extent of digestion.
In the first example the temperature was decreased from 60°C to 37°C. As a result the digestion was reduced from 99% to 70% of the total protein injected on the enzyme column.

Peptide digest from Cytochrome c separated on a
STYROSä
2R/XH 4.6 X 250 mm.
Digestion flow rate 0.15 ml/min. Temperature 37
°C.
By increasing the flow rate from 0.15ml/min to 0.2ml/min during digestion, only 80% of the injected protein is digested.

Peptide digest from Cytochrome c separated on a
STYROSä
2R/XH 4.6 X 250 mm.
Digestion flow rate 0.2 ml/min. Temperature 60
°C.