APPLICATION NOTEAN102201-19 Resolving Efficiency of Longer Columns: case application with medium capacity cation exchanger. Chromatographic columns with a wide range of mechanical and chemical stability provide the possibility of being used within a wide range of conditions without any restrictions. When column length remains the only parameter to achieve resolution, hard gel chromatographic media such as STYROS™ offer an additional advantage over soft gel. The following examples show the separation of 4 proteins at different flow rates and the effect of column length on resolution. The columns are packed with STYROS™ SE, a medium capacity strong cation exchanger and run at two different flow rates.
Table 1. Operating Parameters.
Although preliminary separations could be achieved at high linear velocities with both the 3.3 and 25cm columns, the longer column displays a clear shoulder at 2.4 min retention time that is not present in the separations with the 3.3 cm column. Further reduction of the speed, from 1,800 to 360 cm/hr, reveals a clear peak at 12.9 min during the separation with the 25cm column. The smaller column would still not be able to resolve the isoform of the protein.
|