![]() |
SIMULATED MONOLITH™ POLYMERIC STATIONARY PHASES FOR LIQUID CHROMATOGRAPHY | |
![]() |
![]() |
|
Separation of Angiotensin Variants at Basic pH.
It is well established that the separations of Angiotensins II and III is not possible on reversed-phase stationary phases at acidic pH.
In order to achieve such separation, which also holds true for numerous other peptides of similar structure, it is necessary to operate at relatively high pH (>10) where the use of silica based media is not possible.
The present application note provides an example where the use of STYROSä becomes a convenient tool to operate at high pH and resolve the separation.
The optimized pore structure of the polymeric matrix that makes up the core of STYROSä 1 and STYROSä 2 can not only operate in full pH range, it also makes it possible to maintain high resolution even at high speed. The following chromatograms highlight such characteristic.

Chromatogram 2.
Separation of Angiotensin variants at pH=11.2, on
STYROSä
1 R/XH at 900 cm/hr.

Chromatogram 2.
Separation of Angiotensin variants at pH=11.2, on
STYROSä
1 R/XH at1,800 cm/hr.
Table 1. Operating Parameters for Chromatograms 1 and 2.
|
HPLC System |
Hewlett Packard 1050 | ||
|
Detector |
214 nm |
||
|
Column |
STYROSä 1 R/XH 100x4.6mm | ||
|
Mobile Phase |
A: 10 mM Phosphate in water, pH=11.2 B: Acetonitrile/water |
||
|
Gradient |
0-40% B in 25 ml | ||
|
Flow rate |
2.5 and 5ml/min |
||
|
Temperature |
Ambient |
||
|
Injection volume |
10
ml |
||
|
Sample: 1 mg/ml each |
1: Angiotensin II, 2: Angiotensin III, 3: Angiotensin I |


Table 2. Operating Parameters for Chromatograms 3 and 4.
|
HPLC System |
Hewlett Packard 1050 | ||
|
Detector |
214 nm |
||
|
Column |
STYROSä 2 R/XH 100x4.6mm | ||
|
Mobile Phase |
A: 10 mM Phosphate in water, pH=11.2 B: Acetonitrile/water |
||
|
Gradient |
0-40% B in 25 ml | ||
|
Flow rate |
2.5 and 5ml/min |
||
|
Temperature |
Ambient |
||
|
Injection volume |
10
ml |
||
|
Sample: 1 mg/ml each |
1: Angiotensin II, 2: Angiotensin III, 3: Angiotensin I |
