SIMULATED MONOLITH™ POLYMERIC STATIONARY PHASES FOR LIQUID CHROMATOGRAPHY | |
![]() |
![]() |
|
AN080397-03
Comparison of STYROS™ 1R/XP and 1R/XH.
The
STYROS™
chromatographic stationary phase family of products has been
developed to permit a seamless transition from analytical to
preparative scale.
In the past, researchers had to make several
compromises concerning pressure drop and efficiency.
Small particle sizes were used for analytical
separations, in order to maximize the theoretical plates. This
resulted in considerable back pressure.
Due to the hardware limitations and cost
considerations, large-size particles have been used in micro
preparative and preparative scale applications, resulting in low
back pressure at the expense of significant loss in theoretical
plates as compared with analytical columns.
We have recognized at
OraChrom, the need for a chromatographic media that provides
both the high performance and the low back pressure at the same
time. Due to our carefully optimized polymerization technique, the
special particle size distribution and the high pressure tolerance
of
STYROS™
media, it is now possible to provide packed columns for high
performance analytical as well as preparative applications without
necessarily sacrificing the efficiency.
The
STYROS™
XP series have been developed to provide high resolution
separations with minimal back pressure.
The XH series, with over 35,000 plates/m, are
intended more as analytical columns for fast separations and optimal
back pressure.
The N/DP
ratio (N: theoretical plates,
DP: pressure drop across the column) is an important
factor that has great significance in determining the practicality
of an LC column as a component of a large scale preparative process.
![]() |
Chromatogram 1. |
![]() |
Chromatogram 2. |
The chromatographic conditions are gathered in the following table:
Table 1. Operating Parameters For Chromatograms 1-4.
|
HPLC System |
Hewlett Packard 1050 | |||
|
Detector |
214 nm |
|||
|
Column |
STYROSÔ 2 R/XP 50x4.6 mm ( Chromatogram 3 & 4 ) |
|||
|
Mobile
Phase |
A: 0.1 % TFA in water B: 0.1 % TFA in Acetonitrile/water (95/5) |
|||
|
Gradient |
15-80% B in 10 ml | |||
|
Flow rate |
1 and 5 ml/min | |||
|
Temperature |
Ambient |
|||
|
Injection volume |
10
ml |
|||
|
Sample: 1 mg/ml each |
1:
Cytochrome C, 2: Lysozyme 3: b-Lactoglobulin, 4: Ovalbumin |
![]() |
Chromatogram 3.
Standard Protein Separation on STYROS™ 1 R/XP at 1 ml/min (360 cm/hr). |
||
![]() |
Chromatogram 4.
Standard Protein Separation on STYROS™ 1 R/XP at 5 ml/min (1,800 cm/hr). |