Fast Protein Separation on STYROSä 1R and 2R.

STYROSä 1 R and 2 R/XH series were designed for fast, high-performance separations of biomolecules.

Both media can typically be run at linear flow rates of 1800 cm/hr and separate mixtures of several proteins.

Chromatograms 1 and 2 provide the separation of 4 proteins: Cytochrome C, Lysozyme, b-Lactoglobulin and Ovalbumin in less than one minute, at room temperature.

Each of the two generations has a specific retentivity that differentiates it from the other. In addition to high selectivity, each stationary phase offers a characteristic protein capacity that provides the end user with more selection.

The design of an optimized pore structure allows full access for large molecules such as proteins, to the inner matrix, making it possible to run the separations at high flow rates without loss of capacity or resolution

Table 1. HPLC Operating Parameters for Chromatograms 1 and 2.

  HPLC System Hewlett Packard 1050    
  Detector 214 nm    
  Column STYROSÔ 1 R/XH 50x4.6mm ( Chromatogram 1 )
STYROSÔ 2 R/XH 50x4.6 mm ( Chromatogram 2 )
   
  Mobile Phase A: 0.1 % TFA in water
B: 0.1 % TFA in Acetonitrile/water (95/5)
   
  Gradient 15-80% B in 1 minutes    
Flow rate 5 ml/min (1,807 cm/hr)
  Temperature Ambient    
  Injection volume 10 ml    
  Sample:
1 mg/ml each
1: Cytochrome C, 2: Lysozyme
3:
b-Lactoglobulin, 4: Ovalbumin
   

Chromatogram 1: Standard Protein Separation on STYROSÔ 1 R/XH

Chromatogram 2: Standard Protein Separation on STYROS Ô 2 R/XH