SIMULATED MONOLITH™ POLYMERIC STATIONARY PHASES FOR LIQUID CHROMATOGRAPHY | |
![]() |
![]() |
|
Fast
Protein Separation on
STYROSä
1R
and
2R.
STYROSä
1 R and
2 R/XH series were designed for fast, high-performance
separations of biomolecules.
Both media can typically be run at linear flow
rates of 1800 cm/hr and separate mixtures of several proteins.
Chromatograms 1 and 2 provide the separation of
4 proteins: Cytochrome C, Lysozyme,
b-Lactoglobulin and Ovalbumin in less than one minute,
at room temperature.
Each of the two generations has a specific
retentivity that differentiates it from the other. In addition to
high selectivity, each stationary phase offers a characteristic
protein capacity that provides the end user with more selection.
Table 1. HPLC Operating Parameters for Chromatograms 1 and 2.
|
HPLC System |
Hewlett Packard 1050 | |||
|
Detector |
214 nm |
|||
|
Column |
STYROSÔ 2 R/XH 50x4.6 mm ( Chromatogram 2 ) |
|||
|
Mobile
Phase |
A: 0.1 % TFA in water B: 0.1 % TFA in Acetonitrile/water (95/5) |
|||
|
Gradient |
15-80% B in
1 minutes |
|||
|
Flow rate |
5 ml/min |
|||
|
Temperature |
Ambient |
|||
|
Injection volume |
10
ml |
|||
|
Sample: 1 mg/ml each |
1:
Cytochrome C, 2: Lysozyme 3: b-Lactoglobulin, 4: Ovalbumin |

Chromatogram 1: Standard Protein Separation on
STYROSÔ
1 R/XH

Chromatogram 2: Standard Protein Separation on
STYROS
Ô
2 R/XH